Advertisement

5.
Repeat
step 3
6.
Enter the number of standards in the run, and then press the Continue screen key.
7.
Enter the location of the first standard vial, and then press the Continue screen key.
Requirement:
8.
Make sure that the solvents listed in the dialog box are primed and in the appropriate inlet
tubes, and then press the Start screen key.
5.1.4

Modify a sample set during a run

You can modify operating conditions during a standalone mode run as follows:
Via the stat run function
By editing the separation method
5.1.4.1
Perform a stat run
You can use the stat run function to interrupt a run in progress and perform one or more injections
after the current injection is done, then continue with the scheduled run. Performing a stat run is
most useful when running separation methods that use identical solvents and temperatures.
To perform a stat run while a sample set is running:
1.
Load the sample(s) into the carousel, and then place the carousel in the sample
compartment.
2.
In the Main screen, press the Run Samples screen key.
3.
Select the separation method or sample set you want to use for the stat run, and then press
the Run screen key.
4.
Enter the carousel position of the vial(s) and other parameters in the appropriate fields in
the screen.
5.
Press the Start screen key.
Result:
A list of the solvents required for the run appears.
6.
Confirm that you prepared the separations module with the indicated solvent(s).
7.
Press the Stat screen key.
Result:
The stat run starts immediately after the current injection is complete. The original
sample set resumes after the stat run ends.
5.1.4.2
Edit a separation method
You can modify a separation method while a sample set is running. If the separation method is
being used for the current injection, the current injection is unaffected. You must save the changes
to the modified separation method before you can apply the method to all subsequent injections
that use that separation method. Follow the procedures on
separation method.
and
step 4
when the sample template includes multiple separation methods.
You must use consecutive standard vials.
February 27, 2018, 715003794 Rev. C
Page 85