Waters Alliance e2695 Operator's Manual page 212

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Table B–7: RS-232 output parameters displayed in the configuration screen
Parameter
Device
Baud rate
B.3.2.2
IEEE-488 connections
Use the IEEE-488 interface connector to connect the separations module to these IEEE-488
devices: (See
figure "Rear panel of the separations module:" on page
Waters 2487, 486, 2414, 2410, and 410 detectors
BusLAC/E card in an Empower software system, a network LAC/E module, or a LAC/E32
Acquisition Server (Empower software)
General purpose instrument board of a MassLynx computer
Requirements:
You cannot use the separations module's IEEE-488 interface at the same time you are
using its Ethernet port for communications.
For remote control, you can only configure the separations module for IEEE-488
communications if it belongs to one of these configurations:
Where the separations module and detector(s) communicate with the data system via
IEEE-488
Where the separations module and data system communicate via IEEE-488, and
include one or more detectors that can only communicate with the data system via
Ethernet (for example, Waters 2998, 2489, 2475, 2424, e-SAT/IN module-connected,
or 3100 Mass Spectrometer Detector.
configurations" on page
The IEEE-488 connector in the separations module mates with a standard IEEE-488 cable.
To avoid damaging components, power-off all instruments connected to the
IEEE-488 bus before you connect an IEEE-488 interface cable to an instrument.
See
figure "Digital signal control of an HPLC system:" on page 35
(IEEE-488 and RS-232) in a typical HPLC system where the separations module functions as the
system controller.
used when the separations module functions under Empower software control.
B.3.2.3
Set the IEEE-488 address
From the Configuration screen, you can view or change the IEEE-488 address of the separations
module. (See
"Configure the operating mode" on page
37.)
See "Empower control" on page 37
February 27, 2018, 715003794 Rev. C
Page 212
Value
None, 746 Integrator, Serial Printer, BCD output
1200, 2400, 9600
48.)
See "Supported IEEE-488 and Ethernet
for the signal connections
for a description of the signal connections
96.)