Waters Alliance e2695 Operator's Manual page 105

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

To enter parameters in the I/O screen:
1.
Press the Next or Prev screen key to display the I/O screen.
Figure 6–9: I/O screen
2.
Enter values in the I/O screen, as appropriate.
The following table describes the parameters in the I/O screen.
Table 6–8: I/O parameters
Parameter
Initial States
Chart Output
Timed Table (screen key)
Function
Defines the initial condition for
each of the four event
switches. At the beginning of
each injection cycle, each
switch returns to the state
defined in this parameter.
Defines the signal sent out on
the chart out terminals on the
rear panel of the separations
module.
Displays the I/O Events table.
February 27, 2018, 715003794 Rev. C
Page 105
Value range
On
Off
Toggle
(Toggle is a one-time change
of state.)
Pulse
(Pulse is a single pulse with a
width defined in the Param
column of the I/O Events
table.)
No Change
Flow Rate
System Pressure
Sample Loop Pressure
%A, %B, %C, %D
Column Temp
Sample Temp
Degasser Vacuum
Primary Head
Pressure
See
"Edit the I/O Events table"
on page
106.