Site Requirements - Waters Alliance e2695 Operator's Manual

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

2.2

Site requirements

The installation site must meet the specifications in the following table.
Table 2–1: Installation site requirements
Factor
Temperature
Relative humidity
Bench space
Vibration
Clearance
Static electricity
Power
Electromagnetic fields
Notice:
To avoid overheating the separations module, provide at least 16 cm (6 inches) of
clearance at the rear of the separations module.
Typical system dimensions for the e2695 Separations Module:
Note:
To avoid exceeding the sizing limits of the tray, use a combination of four 1-L bottles or two
4-L bottles plus two 1-L bottles.
The top panel of the separations module can support the weight of two detectors, the detector drip
tray, and the solvent bottle tray with four filled solvent bottles. Be sure to provide leak protection
below any fluid-carrying units you place on top of the separations module.
Requirement
4 to 40 °C (39 to 104 °F)
20 to 80%, noncondensing
Width: 45.7 cm (18 inches). With optional
column heater or column heater/cooler, 58.4
cm (23 inches).
Depth: 57.1 cm (22.5 inches). With optional
sample heater/cooler, 64.8 cm (25.5 inches)
Height: 57.1 cm (22.5 inches)
Level to within ±2°
Negligible
At least 16 cm (6 inches) at rear for ventilation
Negligible
Grounded AC, 100-240 Vac, 50-60 Hz
No nearby source of electromagnetic noise
such as arcing relays or electric motors
February 27, 2018, 715003794 Rev. C
Page 38