Create And Edit A Sample Set - Waters Alliance e2695 Operator's Manual

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

The following table describes the actions that you can specify with Waters 2487 and 486 tunable
absorbance detectors.
Table 6–16: 2487 and 486 detector actions
Parameter
Set wavelength
Set filter
Set AU full scale
Set chart mark
Set chart polarity
Auto zero
Set lamp
6.3

Create and edit a sample set

A sample set is a table of instructions that the separations module uses to perform an automated
run in a standalone mode. During a run, the separations module performs the function in each row
of the table sequentially.
To create and edit a sample set:
1.
On the Main screen, click the Develop Methods screen key.
2.
On the Methods screen, click the New screen key.
3.
Click the Sample Set screen key.
4.
Name the sample set, and then click Enter.
Tip:
The separations module automatically adds the appropriate extension.
5.
Enter information in the Sample Set table as appropriate.
6.
Press Exit to display a dialog box that asks if you want to save the changes.
Function
Sets the wavelength of the
detector.
Sets the time constant of the
filter in the detector.
Sets the absorbance units full
scale to adjust the sensitivity
of the detector.
Sends a chart mark signal to
the output terminals.
Sets the polarity of the analog
output.
Performs an autozero on the
detector.
Turns the lamp on and off.
February 27, 2018, 715003794 Rev. C
Page 113
Value range
2487: 190 to 700 in 1 nm
increments
486: 190 to 600 in
1 nm increments
2487: 0.1 to 99.0 in 0.1 sec
increments
486: 0.1 to 5.0 in 0.1 sec
increments
2487: 0.0001 to 4.0000 in
0.0001 AU increments
486: 0.001 to 2.000 in 0.001
AU increments
2487 and 486: (No values)
2487 and 486: +, –
2487 and 486:
(No values)
2487 and 486: On, Off