7.6.4

Clean the sample compartment

Clean the sample compartment when these conditions apply:
The sample carousels do not turn properly.
The carousel carrier does not engage.
Required materials
Wrench, hex, 3-mm
Nonabrasive cleaner
To access the sample compartment:
1.
Open the sample compartment and syringe cover doors (see
system components (front view)" on page
2.
Remove the five carousels.
Tip:
Use the Door is Open dialog box to move the carousel carrier to the "A" position (see
"Load carousels" on page
3.
Open the solvent delivery tray and solvent conditioning tray access doors (see
"Access the sample compartment" on page
4.
Use the hex wrench to loosen the two sample compartment front panel screws.
Figure 7–16: Access the sample compartment
Solvent
delivery tray
access door
5.
Pull the sample compartment front panel towards you to remove the panel.
133).
62).
147).
Sample compartment
front panel screws
February 27, 2018, 715003794 Rev. C
Page 147
figure "Sample management
figure
Solvent
conditioning tray
access door