Table 6–18: Sample Set functions (continued)
Function name
System Prep
Purge
Equil
Condition
Auto Add
1. For typical unattended operation, leave the "Hold when time expires" check box (in the Equilibrate dialog box)
unselected. Select the check box and set a time only if you want to maintain (hold) initial conditions once the set time
expires (and respond to the two soft keys, "Resume Sample" and "Abort Samples", that appear in the Status screen).
Pressing one of these soft keys performs the desired action. Note that elapsed time continues until you press one of the
keys.
6.3.0.1
Auto Standards
Use the AutoStds function to repetitively run standards after a specified number of vials or
injections. You can specify multiple standards vials to support large runs.
To use the Auto Standards function:
1.
In the Sample Set screen, select AutoStds from the drop-down list of functions, and then
press the Row Details screen key to display the Auto Standards dialog box.
2.
Enter the frequency with which you want to run standards, and then select either vials or
injections.
3.
Enter the location of each standard vial and its capacity in the Source Vial(s) table.
When you start the run, the separations module runs the standard. When the current
standards vial is empty (based on the number of injections, the size of the injections, and
the total volume), the separations module automatically advances to the next standards vial
specified in the Source Vial(s) table.
Description
Performs the System Prep function using the
parameters from the separation method. For
more information on System Prep, see
System Prep" on page
Purges the sample management system.
Press the Row Details screen key to enter
additional purge parameters.
Delivers solvent through the system for the
specified period of time using the initial
conditions in the specified separation method.
Press the Row Details screen key to display
the Equilibrate dialog box
Equil parameters.
Delivers to the column the gradient specified in
the separation method; conditions the column
without injecting samples.
Injects sample from vials in the selected order.
See
February 27, 2018, 715003794 Rev. C
Page 115
75.
1
to enter additional
"Auto Add" on page
116.
"Run