Set Column Parameter Values - Waters Alliance e2695 Operator's Manual

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

6.2.4

Set column parameter values

To enter parameters in the Column screen:
1.
To display the Column screen, press Next or Prev.
Figure 6–8: Column screen
2.
Enter values in the Column screen, as appropriate.
The following table describes the parameters in the Column screen.
Table 6–7: Column parameters
Parameter
Column Temperature Target
On error
Column Temperature Range
Function
Sets the temperature of the
column heater or column
heater/cooler (if installed). To
turn off the column heater or
column heater/cooler, press
the Clear key.
Response that occurs when
column temperature is out of
the specified range.
Sets the maximum allowable
variation in column
temperature. If the
temperature variation exceeds
the range, the alarm condition
selected in the adjacent box is
triggered.
February 27, 2018, 715003794 Rev. C
Page 103
Value range
Column heater:
20 to 65 °C in 1 °C increments
(5 °C above ambient)
Column heater/cooler:
4 to 65 °C
(the greater of 15 °C below
ambient or 4 to 65 °C)
See
table "Alarm responses"
on page
97.
±10 °C in 1 °C increments

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents