Modify a sample set during a run

You can modify operating conditions during a standalone mode run as follows:
Via the stat run function
By editing the separation method
Perform a stat run
You can use the stat run function to interrupt a run in progress and perform
one or more injections after the current injection is done, then continue with
the scheduled run. Performing a stat run is most useful when running
separation methods that use identical solvents and temperatures.
To perform a stat run while a sample set is running:
1.
Load the sample(s) into the carousel, and then place the carousel in the
sample compartment.
2.
In the Main screen, press the Run Samples screen key.
3.
Select the separation method or sample set you want to use for the stat
run, and then press the Run screen key.
4.
Enter the carousel position of the vial(s) and other parameters in the
appropriate fields in the screen.
5.
Press the Start screen key.
Result:
6.
Confirm that you prepared the separations module with the indicated
solvent(s).
7.
Press the Stat screen key.
Result:
complete. The original sample set resumes after the stat run ends.
You must use consecutive standard vials.
A list of the solvents required for the run appears.
The stat run starts immediately after the current injection is
Perform automatic runs in a standalone mode