Table C–3: Electrical specifications (continued)
Attribute
Hold inject (input)
Chart output
Run stopped
Inject start
Ground terminals
Specification
Two sets of (+, -) terminals that allow other LC devices to delay an
injection
Boolean operator selects whether one or both inputs delay the
injection
Input voltage range: ±30 Vdc
Logic high = > 3.0 Vdc ±10%, logic low = < 1.9 Vdc ±10%
Minimum pulse width = 10 msec
Two terminals (+, -) for recording the following user-selectable
outputs:
•
Programmed flow rate
•
Sample loop pressure
•
System pressure
•
Primary head pressure
•
Programmed composition (%A, %B, %C, %D)
•
Sample temperature
•
Column temperature
•
Degasser vacuum
Contact closure indicating:
•
Sample set is suspended
•
Current function is suspended
Maximum allowable current = 0.5 A
Maximum allowable voltage = 30 Vdc
Contact resistance = 0.2 ohms
Contact closure (across terminals 1 and 2) for 1 second on injection
Maximum allowable current = 0.5A
Maximum allowable voltage = 30 Vdc
Contact resistance = 0.2 ohms
Connected to signal ground and used as reference for
outputs
February 27, 2018, 715003794 Rev. C
Page 219