Waters Alliance e2695 Operator's Manual page 209

Separations module
Hide thumbs Also See for Alliance e2695:
Table of Contents

Advertisement

Figure B–13: I/O signal connectors
Connector B
B.3.1.1
I/O signals
The following table describes the signals sent via the I/O connectors of the separations module.
See
Appendix C
Table B–4: I/O Signals
Signal
Inject start
Stop flow
Hold inject 1 and 2
1
Inject Start
2
3
Ground
4+
Stop Flow
5-
6+
Hold Inject 1
7-
8+
Hold Inject 2
9-
Ground
10
11
Chart Out
12
for information about the electrical specifications of each signal.
Description
An output (contact closure) to indicate that an injection began.
Connect a wire pair to the two Inject start terminals.
An input that enables other LC components to stop the solvent flow from
the separations module. You specify the signal conditions that stop the
flow using the Events In screen key in the Configuration screen.
Connect the positive input wire to the "Stop Flow +" terminal and the
negative input wire to the "Stop Flow -" terminal.
Inputs that enable other LC components to delay the injection (for
example, an external autosampler used in the Operate Gradient by Event
In mode or a fraction collector). You can specify the signal conditions
(using the Events In screen key in the Configuration screen) that prevent
the separations module from making an injection.
Connect the positive input wire to the appropriate "Hold Inject +" terminal
and the negative input wire to the appropriate "Hold Inject -" terminal.
February 27, 2018, 715003794 Rev. C
Connector A
1
Switch 1
2
3
Ground
4
Switch 2
5
6
Switch 3
7
8
Switch 4
9
Ground
10
11
Run Stopped
12
Page 209