5.1.5

Stop a run

To stop a run in progress, press one of the following keys from the Status screen:
Stop Inject (screen key) – Stops the current injection, and displays a dialog box requesting
you to abort or resume the injection.
Hold Inject (screen key) – Completes the current function, suspends operation of
subsequent functions, and displays screen keys that allow you to abort or resume the
sample set.
Hold Gradient (screen key) – Holds the current gradient conditions. A screen key allows you
to resume the gradient.
Alternative:
On the keypad, press Stop Flow to halt the flow of solvent, suspend operation of the
current function, and display screen keys that allow you to abort or resume the function.
If any alarm condition specified in the separation method, sample set, or sample template is set to
"Stop Function" or "Stop Flow" when alarm condition occurs, a dialog box appears notifying you of
the alarm condition. The following screen keys are in this dialog box:
Abort – Cancels the current function
Resume – Resumes the current function
5.2
Perform automatic runs under Empower software
control
If your separations module is configured for Empower Chromatography Data Software, you make
an automatic run from the Empower software Run Samples window. When the separations
module is running under Empower software control, the word "Remote" appears in the banner
area.
Configure your separations module for Empower software control as described in
operating mode" on page
To make an automatic run, perform the following steps at the Empower workstation:
Tip:
See the Empower online help for an explanation of how to perform the tasks in these steps.
1.
Depending on the type and version of remote control software you are using, create a
chromatographic system as follows:
For Empower software – Create a system that includes the separations module.
2.
Create an instrument method that uses the system you defined in
Tip:
Any parameter values not explicitly set in the Empower instrument method
automatically adopt the values in the default separation method in the separations module.
These values include:
Column selection valves (3-column, 6-column, and 2-column regeneration valves)
recognized by the e2695 software
A needle- wash time selection of "normal", "double", and "extended"
52.
February 27, 2018, 715003794 Rev. C