Figure 7–7: Replacing the in-line filter
Inlet housing
6.
Reconnect the in-line filter inlet and outlet housings.
7.
Re-tighten the compression screw into the inlet housing.
8.
Flush the solvent management system with mobile phase at 1 mL/min for 10 minutes (see
"Power-off the separations module" on page
9.
Inspect the in-line filter for leaks, and tighten the fittings as required.
In-line filter
element
64).
February 27, 2018, 715003794 Rev. C
Page 132
Outlet housing