Corporation assumes no responsibility for any errors that may appear in this document. This document is believed to be complete and accurate at the time of publication. In no event shall Waters Corporation be liable for incidental or consequential damages in connection with, or arising from, its use.
Contacting Waters ® Contact Waters with enhancement requests or technical questions regarding the use, transportation, removal, or disposal of any Waters product. You can reach us via the Internet, telephone, or conventional mail. Waters contact information Contacting medium Information...
United States and Canadian safety requirements Consult instructions for use Audience and purpose This document describes the installation of the e-SAT/IN module. It is intended is intended to be used by qualified operators, installation engineers, and field service engineers.
Intended use of the Waters e-SAT/IN module Waters designed the e-SAT/IN module to translate analog signals to digital signals for data collection. You can operate the device with instruments used to analyze various types of compounds. The e-SAT/IN module is for research use only. Calibrating To calibrate LC systems, follow acceptable calibration methods using at least five standards to generate a standard curve.
Safety advisories ....................iv Operating this instrument ................. iv Applicable symbols ..................... iv Audience and purpose..................iv Intended use of the Waters e-SAT/IN module............ v Calibrating ......................v Quality-control ..................... v ISM classification ....................v ISM Classification: ISM Group 1 Class B ............v EC authorized representative ................
Page 8
Ethernet connections ..................2-3 Single e-SAT/IN module connection ............... 2-3 Multiple e-SAT/IN connections ............... 2-4 Network installation guidelines..............2-4 Integrating an e-SAT/IN module to an existing busSAT/IN system..... 2-6 Connecting analog cables ................2-7 Connecting event cables .................. 2-8 Event in connections..................2-8 Event out connections..................
Page 9
Warning symbols ....................A-2 Task-specific hazard warnings................ A-2 Specific warnings ..................... A-3 Caution symbol ....................A-5 Warnings that apply to all Waters instruments ......... A-6 Electrical and handling symbols ..............A-11 Electrical symbols ..................A-11 Handling symbols ..................A-12 B Specifications ..................B-1 C Spare Parts ....................
(Empower™) can collect data from either a single channel or both analog channels. The data start and stop times are identical for both channels. The e-SAT/IN module uses a dedicated local area network (LAN) to communicate to the chromatography data acquisition computer. You can control the e-SAT/IN module using Waters Empower software.
Data status indicator (LED). • Dedicated external voltage autosensing power supply. Features This section describes the features of the e-SAT/IN module. Baseline noise As analysis times decrease, higher data sampling rates are required to accommodate narrower peaks. The e-SAT/IN module features new technology and a design that reduces chromatographic baseline noise at all its sampling rates.
12.5 Ethernet I/O connection The Waters e-SAT/IN module connects to a chromatography data system using a 10 base-T Ethernet interface to a dedicated local area network (LAN). Ethernet allows for the use of long cables (when required) and eliminates the need for proprietary interface hardware beyond the dedicated Ethernet card (network interface card) within the acquisition computer.
24-bits of A/D conversion (depending on the sampling rate). Analog input channels The analog input channels of the e-SAT/IN module receive analog signals from the detector. This analog signal is converted to digital data. The digital data signal is bundled into a data packet and sent using Ethernet to the acquisition computer as shown in the figure below.
3-2 for indicator details. IP address sequence The green LED indicator shows that the e-SAT/IN module is requesting IP configuration information by blinking. After the module receives its IP configuration information, the LED remains on and the e-SAT/IN module is ready for use.
External power supply The Waters e-SAT/IN module uses an external power supply that is provided with the device. The power supply automatically adjusts to the AC input line voltage, and provides the proper DC voltage to the e-SAT/IN module. A separate AC power supply cable is provided for the country in which it will be installed.
Installing the Module This chapter provides information you need to install a single e-SAT/IN module or a multiple e-SAT/IN system. Contents Topic Page Selecting the site Making electrical connections Installing the network interface card Connecting analog cables Connecting event cables...
Do not install the e-SAT/IN module and the associated network components where they can be subjected to fluidic leaks. • Do not install the e-SAT/IN module close to devices that emit high levels of electromagnetic interference (EMI) or radio frequency interference (RFI).
(NIC) printed circuit board (customer supplied) provides the interface that allows communication to occur. You must install the e-SAT/IN module software driver in the acquisition computer before the computer will recognize the presence of the e-SAT/IN module.
The Ethernet switch allows you to connect multiple e-SAT/IN modules to the acquisition computer. The acquisition computer identifies each e-SAT/IN module by its network address.
Page 23
Multiple e-SAT/IN Ethernet connections Network interface card Acquisition computer 10 base-T Ethernet cable Output Ethernet switch 10 base-T Input connectors Ethernet c able (one per e-SAT/IN) e-SAT/IN e-SAT/IN e-SAT/IN e-SAT/IN Analog cables Detectors Installing the network interface card...
Integrating an e-SAT/IN module to an existing busSAT/IN system The e-SAT/IN module can be integrated to an existing busSAT/IN configuration. You must connect the e-SAT/IN module (or Ethernet switch for multiple e-SAT/IN module configurations) to a network interface card (NIC) (customer supplied) within the acquisition computer.
Connecting analog cables Analog cables connect the detectors to the e-SAT/IN module. The front panel of the e-SAT/IN module provides a separate connector for the two analog input channels. Refer to the operating instructions supplied with the detector for specific connection details.
Event in connections The e-SAT/IN module can accept either a TTL switch closure or pulsed inject start signal from a manual injector or autoinjector. The inject start signal triggers the e-SAT/IN module to start collecting and converting data.
Page 27
1. Because both Channel 1 and 2 share the same timebase, both channels will be triggered at the same time. To connect event cables: Remove the event connector from the e-SAT/IN module. Connect the two spade connectors at one end of an Event In/Out cable to: •...
To injector Connecting the BCD cable The BCD (binary coded decimal) connection on the e-SAT/IN module allows the module to read and record BCD-encoded data from your controller instrument, such as an autosampler, that is compatible with the supported BCD data format listed in the table titled “BCD connection assignments”...
− Least significant bit of most significant nibble Connecting the power supply To make the electrical connections for the e-SAT/IN module(s): Ensure that the power switch on the front of each module is set to the Off (0) position. Connecting the power supply...
Page 30
Connect the power supply cable to the e-SAT/IN module before connecting it to the grounded AC source. Connect the power supply to a properly grounded AC source. The power supply provided specifically for the e-SAT/IN module is voltage autosensing, and will adjust accordingly within ratings listed on the device label.
Using the e-SAT/IN Module This chapter describes the recommended power-on procedure for the e-SAT/IN module. It also discusses how to configure the e-SAT/IN module from Empower software, as well as how to troubleshoot the e-SAT/IN module. Perform all procedures in the order specified in this chapter.
Refer to the specific guides for each device for power-on procedures. The e-SAT/IN module has a power switch on the front panel. The switch is a pushbutton switch that controls the power from the power supply.
You need to access the Instruments tab of the Acquisition Server Properties page of the Configuration Manager, and scan for instruments. Once the e-SAT/IN module is displayed and Ok ? is Yes, then a chromatography system can be created. After creating a system, you can set up or edit the instrument method.
Page 34
Empower Pro interface To configure the e-SAT/IN module using the Instrument Method Editor in the Empower Pro interface: Log in to Empower, then click Browse Project. The Browse Project dialog box opens. Empower Pro window and Browse Project dialog box Select a project, then click OK.
Page 35
Active instrument list When an e-SAT/IN module is listed, you can a modify the preferences using the Instrument Method Editor. When the e-SAT/IN module(s) is not listed, you can add a new instrument to designate the e-SAT/IN module as an active device.
Page 36
Instrument Method Editor The Instrument Method Editor appears. This window has a selection of instrument icons located in the window taskbar. Click e-SAT/IN. The e-SAT/IN General tab appears. Using the e-SAT/IN Module...
Page 37
General tab The General tab contains specific preferences that affect the performance of the e-SAT/IN module. Refer to the “General Tab” topic in the Empower Help. Selecting sampling rates Quality data sampling relies on minimizing extraneous electrical noise in the environment.
Page 38
To configure the e-SAT/IN module for BCD use: Fill in the rows in the Samples table for the required number of injections. Connect the e-SAT/IN module BCD port to the autosampler using a BCD cable. Ensure that the following conditions are defined for the e-SAT/IN module: Specify the e-SAT/IN module as one of your instruments.
Page 39
1 AU equals 1 V, enter a scale factor of 1.0. The default setting is 1000. e-SAT/IN Events tab Use the Events tab to define operation and initiation of the selected e-SAT/IN module output events. Configuring the e-SAT/IN module...
Page 40
Enabling events Enabling an event output generates a contact closure that can initiate a process on another device (at a specified time after the e-SAT/IN event clock starts). The contact closure can control operation of a fraction collector or chart recorder, or activate solvent/column switching valves.
Page 41
The Relay Event of the e-SAT/IN module can function as either a standard programmable event or as part of the ready loop, based on your selection in the Ready Operating Mode area.
Page 42
Comments − Optional description of the programmed event. The description can be up to 256 characters. After the e-SAT/IN instrument method programming is completed and saved, you can run samples. Manually start the autosampler. Data acquisition begins automatically when Empower software receives an inject signal from the autosampler.
Page 43
Accessing QuickStart Empower QuickStart opens the Run Samples window. Configuring the e-SAT/IN module 3-13...
Page 44
Run Samples window Access the instrument lists by selecting View > Method > Instrument. View Instrument Method window Instrument selection control 3-14 Using the e-SAT/IN Module...
When an e-SAT/IN module is listed, you can modify the preferences using the Instrument Method Editor. When the e-SAT/IN module(s) is not listed, you can add a new instrument to activate the e-SAT/IN module as an active device. The Instrument Method Editor appears. This window has a selection of instrument icons located in the window header.
Page 46
Save As command. View the smoothing derived channel in Review, as follows: In the Project window Injections Table view, select the sample or standard injection to which you want to apply smoothing. Access Review. 3-16 Using the e-SAT/IN Module...
Select Process. Select Apply Method Set and specify the name. Troubleshooting During power-on and operation, the e-SAT/IN module indicates its status using the LEDs on the front panel. Refer to the following table for a summary of the LED indicators.
• Power supply cable is properly attached to the e-SAT/IN module. • e-SAT/IN module power switch is set to the on (1) position. System troubleshooting To perform basic system troubleshooting: Examine the system, checking the simple things first. Is something...
When to call Waters technical service Many problems with an e-SAT/IN module can be easily corrected. However, if you cannot correct a condition, contact Waters Technical Service at 800 252-4752, U.S. and Canadian customers only. Other customers, call your local Waters subsidiary or Technical Service Representative, or call Waters corporate headquarters in Milford, Massachusetts (U.S.A.).
Error messages Error messages related to the e-SAT/IN module will appear in the Empower Message Center Log. Click the Message Center icon in the taskbar to access the log. Error messages Error number Description This occurs if the internal buffer in the e-SAT/IN module is filled.
Page 51
Error messages (Continued) Error number Description An invalid number of events argument was specified in the Setup command. An invalid channel argument was specified in the Get Data command. An invalid max points argument was specified in the Get Data command. Troubleshooting 3-21...
Page 53
Safety Advisories Waters instruments display hazard symbols designed to alert you to the hidden dangers of operating and maintaining the instruments. Their corresponding user guides also include the hazard symbols, with accompanying text statements describing the hazards and telling you how to avoid them.
Heed all warnings when you install, repair, and operate Waters instruments. Waters assumes no liability for the failure of those who install, repair, or operate its instruments to comply with any safety precaution.
The following warnings can appear in the user manuals of particular instruments and on labels affixed to them or their component parts. Burst warning This warning applies to Waters instruments fitted with nonmetallic tubing. Pressurized nonmetallic, or polymer, tubing can burst. Warning: Observe these precautions when working around such tubing: •...
Page 56
Also ensure a gas-fail connection is connected to the LC system so that the LC solvent flow stops if the nitrogen supply fails. Mass spectrometer shock hazard This warning applies to all Waters mass spectrometers. To avoid electric shock, do not remove the mass spectrometer’s Warning: protective panels.
Biohazard warning This warning applies to Waters instruments that can be used to process material that might contain biohazards: substances that contain biological agents capable of producing harmful effects in humans. Waters instruments and software can be used to analyze or...
Warnings that apply to all Waters instruments When operating this device, follow standard quality control procedures and the equipment guidelines in this section. Attention: Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
Page 59
• Keine Schläuche verwenden, die stark geknickt oder überbeansprucht sind. • Nichtmetallische Schläuche nicht für Tetrahydrofuran (THF) oder konzentrierte Salpeter- oder Schwefelsäure verwenden. • Durch Methylenchlorid und Dimethylsulfoxid können nichtmetallische Schläuche quellen; dadurch wird der Berstdruck des Schlauches erheblich reduziert. Warnings that apply to all Waters instruments...
Page 60
Attenzione: fare attenzione quando si utilizzano tubi in materiale polimerico sotto pressione: • Indossare sempre occhiali da lavoro protettivi nei pressi di tubi di polimero pressurizzati. • Spegnere tutte le fiamme vive nell'ambiente circostante. • Non utilizzare tubi eccessivamente logorati o piegati. •...
Page 61
농축 질산 또는 황산과 함께 사용하지 마십시오. • 염화 메틸렌(Methylene chloride) 및 디메틸술폭시드(Dimethyl sulfoxide)는 비금속 튜브를 부풀려 튜브의 파열 압력을 크게 감소시킬 수 있으므로 유의하십시오. 警告:圧力のかかったポリマーチューブを扱うときは、注意してください。 • 加圧されたポリマーチューブの付近では、必ず保護メガネを着用してください。 • 近くにある火を消してください。 • 著しく変形した、または折れ曲がったチューブは使用しないでください。 • 非金属チューブには、テトラヒドロフラン(THF)や高濃度の硝酸または硫酸などを流 さないでください。 • 塩化メチレンやジメチルスルホキシドは、非金属チューブの膨張を引き起こす場合が あり、その場合、チューブは極めて低い圧力で破裂します。 Warnings that apply to all Waters instruments...
Page 62
Warning: The user shall be made aware that if the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. Attention: L’utilisateur doit être informé que si le matériel est utilisé d’une façon non spécifiée par le fabricant, la protection assurée par le matériel risque d’être défectueuses.
Electrical and handling symbols Electrical symbols These can appear in instrument user manuals and on the instrument’s front or rear panels. Electrical power on Electrical power off Standby Direct current Alternating current Protective conductor terminal Frame, or chassis, terminal Fuse Recycle symbol: Do not dispose in municipal waste.
Handling symbols These handling symbols and their associated text can appear on labels affixed to the outer packaging of Waters instrument and component shipments. Keep upright! Keep dry! Fragile! Use no hooks! A-12 Safety Advisories...
Specifications This appendix includes specifications for the following: • Physical • Performance • Analog input • BCD input • Power supply • Communication interface Physical specifications Item Specification Dimensions Height = 1.25 inches (3.18 cm) Depth = 6.25 inches (15.9 cm) Width = 5.25 inches (13.3 cm) Weight 530 grams (15.1 ounces)
Page 66
Performance specifications (Continued) Item Specification Voltage accuracy Output within 1% of input Impedance (common mode) 500 K ohms, +/− 1%, minimum Impedance (differential) 2 M ohm, +/− 1%, minimum Nonlinearity 0.01% of full scale Precision +/− 0.1% for inputs greater than 0.5 mV Difference in output +/−...
Page 67
BCD performance specifications (Continued) Item Specification Electrical signal TTL compatible, all 8 bits characteristics • V = 2.0 Volts, maximum • V = 0.8 Volts, minimum • I = 20 μA, maximum • I = 0.1 mA maximum • I = −...
Spare Parts This appendix lists recommended spare parts for the e-SAT/IN module. The e-SAT/IN module and power supply contain no internal user-serviceable parts. The e-SAT/IN module and a replacement is offered as an exchange stand-alone unit only. Spare parts Item...
Page 70
Spare parts (Continued) Item Part number Empower Software Instrument Driver 667001494 Disk Kit Spare Parts...
2-12 Events tab caution symbol events, enabling 3-10 chemical hazard warning communication specifications configuring the e-SAT/IN module flammable solvents connecting BCD cable 2-10 event cables General tab power supply 2-11 guidelines, network installation contacting Waters Technical Service...
Page 72
2-12 installation site selection connecting 2-11 Instrument Method Editor specifications integrating e-SAT/IN module to powering on the e-SAT/IN module busSAT/IN system power-on problems 3-18 intended use power-on self-tests interfaces Pro interface Empower Pro purpose and audience...
Page 73
3-15 spare parts specifications analog channel performance BCD performance communication performance physical power supply summary of LED indicators 3-17 switch, Ethernet symbols caution electrical A-11 handling A-12 warning system troubleshooting 3-18...