Table of Contents

Advertisement

4.
If the detector is a 2424, remove the barbed fitting from the underside of the detector, as this
interferes with the mounting of the drip tray.
5.
Remove the T-10 screw located on the bottom right corner of the detector front panel and
set it aside.
Figure B–3: A- and C-series drip tray mounting location on detector
Drip tray
mounting flange
6.
Position the drip tray so that the mounting hole for the T-10 screw you removed is aligned
between the mounting ears on the right end of the drip tray.
7.
Push the mounting slot on the opposite side of the drip tray into the mounting flange on the
detector until the drip tray locks into place.
8.
Insert the T-10 screw you removed earlier in the hole between the drip tray mounting ears
and secure the drip tray to the detector.
Figure B–4: A- and C-series drip tray installed
9.
Install the elbow fitting on the right end of the drip tray.
February 27, 2018, 715003794 Rev. C
Page 196
T-10 screw
Drip tray mounting
ears (insert T-10
screw here)
Elbow fitting goes
here