Front Panel Control

This chapter explains how to manually control the components of your HPLC system through the
front panel of the separations module.
When the separations module is configured in the No Interaction or Operate Gradient by Event In
mode, you control each HPLC system component individually via its front panel. When the
separations module is configured in the System Controller mode, you control all components in
your HPLC system through the front panel of the separations module.
4.1

Routine start-up

If the separations module is not already powered-on, follow the power-on procedure in
When the start-up diagnostics tests run to completion, the separations module enters the idle
mode.
To automatically identify the reports and methods you will generate, press Enter, then select your
name in the User list. If your name is not on the list, you can enter it by pressing the Config screen
key (to display the Configuration screen), pressing the User Names screen key (to display the
Users dialog box), then entering your name using the keypad. If necessary, refer to
alphanumeric string" on page
The time and date are entered into the log file.
4.1.1

Reinitialize the system

It is good practice to reinitialize the system instruments at least once a week. When the
separations module and detectors are power-cycled (powered-off for a minimum of one minute,
and then powered-on), each undergoes a complete system initialization, including internal
diagnostics and verifications. This procedure helps ensure that the mechanical and electrical
components, as well as internal firmware, synchronize for proper operation.
4.2

Load sample vials

Loading the sample vials involves these tasks:
Removing the carousel from the sample compartment
Loading the vials into the carousel
Loading the carousel into the sample compartment
45. A user name is not required to operate the separations module.
February 27, 2018, 715003794 Rev. C
Page 67