Figure 7–13: Injector assembly components
Injector motor cable
Needle tee
Compression screw
Captive mounting screws
To access the injector assembly:
Note:
This procedure applies only to a separations module with no column heater or column
heater/cooler option installed. If the column heater or column heater/cooler option is installed, see
"Remove side panel of separations module" on page 135
side panel of the separations module.
1.
Power-off the separations module, and disconnect it from the electrical outlet.
2.
Remove the 2 screws that secure the separations module's right-hand side panel.
3.
Slide the side panel to the rear, off the rear panel posts, to access the injector assembly.
February 27, 2018, 715003794 Rev. C
Page 142
Sensor cable
Needle
Upper frit retainer
Seal pack assembly
(exploded view)