4.2.1

Remove the carousel

To remove a carousel from the sample compartment:
1.
Open the carousel door.
2.
Press the Next screen key, or select the desired carousel screen key (A to E), to move the
desired carousel to the front loading position.
3.
Remove the carousel from the sample chamber.
4.2.2

Load vials

Use only 12  32-mm (2-mL) sample vials.
Insert the prepared vials into the appropriate positions in the carousel(s). Be sure to place vials in
the numbered slots that correspond to the sample set (if any) you are using. For an overview of the
vial positions specified in the sample set, display the vial numbers in the Loading View of the
Sample Set table (see
vial positions containing the samples in the Inject Samples dialog box (see
page
79).
4.2.3

Load the carousel

To load the carousel:
1.
Open the carousel door if it is closed.
2.
Press the Next screen key (or select the desired carousel screen key) to position the
carousel turntable for loading the appropriate carousel.
3.
Load the carousel into the sample compartment, and close the carousel door.
Tip:
If you inadvertently load a carousel in the wrong location when the "Verify Carousel
Placement" parameter in the Configuration screen is selected, an error message appears
when you make an injection from that carousel.
4.2.4

Add new sample vials during a run

If you need to add new sample vials while the instrument is processing a sample set, use these
guidelines to minimize interruptions during the run:
Have the new sample vial(s) ready so that you minimize the amount of time that the sample
compartment door is open.
Wait until the run time displayed on the front panel is greater than 30 seconds before adding
the sample(s).
Do not add any sample(s) during the last 30 seconds of the run.
Ensure that the sample compartment door is closed at the start of each injection when
injection commands (selecting vial, washing the needle, drawing sample, and so forth) are
executing.
"View sample sets" on page
February 27, 2018, 715003794 Rev. C
Page 68
83). If you are not using a sample set, enter the
"Inject samples" on