TCP/IP parameter configuration

To operate properly in a network environment, the LAN interface must be
configured with valid TCP/IP network parameters. These parameters are:
• IP address
• Subnet Mask
• Default Gateway
The TCP/IP parameters can be configured by the following methods:
• by automatically requesting the parameters from a network-based DHCP
• by manually setting the parameters using Telnet
• by manually setting the parameters using the Local Controller
The LAN interface differentiates between several initialization modes. The
initialization mode (short form 'init mode') defines how to determine the
active TCP/IP parameters after power-on. The parameters may be derived
non-volatile memory or initialized with known default values. The
initialization mode is selected by the configuration switch, see
page 287.
Agilent InfinityLab LC Series Vialsamplers User Manual
Server (using the so-called Dynamic Host Configuration Protocol). This
mode requires a LAN-onboard Module or a G1369C LAN Interface card, see
"Setup (DHCP)"
on page 290
LAN Configuration

TCP/IP parameter configuration

Table 23
13
on
285