Setting the buzzer volume (BUZZER)
Setting the contrast (CONTRAST)
Setting the discharge mode (DISCH MODE)
5-46
The buzzer volume can be set to 0 through 10 (at a resolution of 1) for FAIL judge-
ment. The buzzer volume for PASS judgement is half that for FAIL judgement.
1.
To display the BUZZER VOL., press the SHIFT +
the SYSTEM2 screen.
2.
Using the F1 key (BUZZER) or the
BUZZER VOL.
To hear the preset sound, press the F1 key (BUZZER).
3.
Using the rotary knob, adjust the buzzer volume.
The LCD screen contrast level can be set to 1 through 10 (at a resolution of 1).
1.
To display the CONTRAST, press the SHIFT +
SYSTEM2 screen.
2.
Using the F2 key (CONTRAST) or the
CONTRAST.
3.
Using the rotary knob, adjust the contrast level.
The contrast level can be set using the SHIFT +
displayed.
You can select the discharge mode to be applied while the output is not turned on.
NORM:
The tester stops discharging when the voltage between the output
terminals fall 30 V or below and the specified discharge time has
elapsed after the test.
CONT:
The tester keeps discharging while the output is not turned on. That
is, the tester keeps discharging even after the specified discharge
time has elapsed.
1.
To display the DISCH MODE, press the SHIFT +
the SYSTEM2 screen.
2.
Using the F3 key (DISCH) or the
MODE.
3.
Using the rotary knob, adjust the discharge mode.
The discharge mode can be set using the SHIFT +
screen displayed.
keys to display
keys, move the cursor to
keys to display the
keys, move the cursor to
keys, regardless of the screen
keys to display
keys, move the cursor to DISCH
keys, regardless of the
TOS9213S/ TOS9213AS