Kikusui TOS9213S Operation Manual page 171

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Program data
Program data
Program data
Program data
Program data
Program data
Program data
Program data
Response
To D:MEM? _<memory number>, the contents of the memory specified by
memory number are returned. The memory contents are returned in the same order
as with program messages.
(Example) If memory 12, at inquiry, contains the memory name DCW12, test volt-
TOS9213S/ TOS9213AS
<timer{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<start voltage>
Data format: Integer
Set value:
0 to 99
Resolution:
1
Unit:
%
<voltage rise time>
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
<WAIT TIME>
Data format: Real number
Set value:
0.3 to 30.0
Resolution:
0.1
Unit:
s
<GND{GUARD|LOW}>
Data format: Character
Set value:
LOW (0), GUARD (1)
<contact check{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<scanner Hi/Low>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
LOW is represented by 0 and HIGH by 1.
<scanner Open>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
OPEN is represented by 1 and NOT OPEN by 0.
OPEN has priority over HIGH/LOW.
age of 5.50 kV, lower reference of 5.00 mA, upper reference of 5.00 mA,
test time of 2.0 s., lower judgment set to ON, timer set to ON, start volt-
age of 20%, voltage rise time of 0.2 s, WAIT TIME of 0.5 s, GND set to
LOW, contact check set to ON, scanners 1 through 5 set to Low, scan-
ners 6 through 10 set to High, and the other scanners set to Open, then,
"DCW12,5.50E3,5.00E-3,5.00E-3,2.0,1,1,20,0.2,
0.5,0,1,992,64512" is returned.
7-47