Kikusui TOS9213S Operation Manual page 67

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Turning the contact check ON/OFF
When the optional high-voltage scanner TOS9221 (with the contact check function)
is connected to the tester, the continuity between the test leadwire and the DUT can
be tested using the HIGH or LOW terminal prior to application of a test voltage. To
do so, turn on the contact check function.
With the high-voltage scanner TOS9220, a check is conducted on the continuity
through to inside of the scanner only.
To turn the contact check function on/off, use the SHIFT + F5 keys.
Each time the SHIFT + F5 keys are pressed, ON and OFF alternate, regardless of
the location of the cursor. The cursor moves to CONTACT CHECK.
Using the
also be made using the rotary knob.
Press the DCW key to return to the DCW1 screen, and then press the START
switch. READY will then disappear from the LCD, and a contact check will start for
each channel.
The test starts as soon as the continuity is confirmed.
The test continues until the test time preset on the timer has elapsed or the STOP
switch is pressed.
The contact check time is calculated using the following formula:
Contact check time = 260 ms + 60 ms x (number of channels set to HIGH or LOW)
When a continuity error is detected in a contact check, " →← FAIL" appears at the
top right of the LCD. On the high-voltage scanner, the LED of the failed channel
lights up in orange.
1.00
TOS9213S/ TOS9213AS
key, move the cursor to the right of CONTACT CHECK. Settings can
Turn the rotary knob clockwise: ON
Turn the rotary knob counterclockwise: OFF
(CH1)
Contact Check 1
(CHn)
TEST
Contact Check n
HV ON
Fig.5-6 Contact check
DCW1
DCW1
LOWER
OFF
UPPER
5.00
kV
mA
VOLT
LOWER
UPPER
→← FAIL
TIMER
OFF
3.0
s
TIMER
5-19