Kikusui TOS9213S Operation Manual page 166

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

SIGPROTECTION (SPR)
SIGPOWERON (SPOW)
SIGCFAIL (SCF)
7-42
Sets PROTECT for Status Signal Output. Also inquires about the current PROTECT
setting.
Message
Setting: SIGPROTECTION_<{ON|OFF}>
SPR_<{ON|OFF}>
SIGPROTECTION?
Query:
SPR?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the PROTECT status of Status Signal Output to OFF,
SPR OFF
Response
To SPR?, the current PROTECT setting for Status Signal Output is returned.
(Example) If the PROTECT status of Status Signal Output is currently set to ON,
1 is returned.
Sets POWER ON for Status Signal Output. Also inquires about the current POWER
ON setting.
Message
Setting: SIGPOWERON_<{ON|OFF}>
SPOW_<{ON|OFF}>
SIGPOWERON?
Query:
SPOW?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the POWER ON status of Status Signal Output to OFF,
SPOW OFF
Response
To SPOW?, the current POWER ON setting for Status Signal Output is returned.
(Example) If the POWER ON status of Status Signal Output is currently set to ON,
1 is returned.
Sets C FAIL for Status Signal Output. Also inquires about the current C FAIL setting.
Message
Setting: SIGCFAIL_<{ON|OFF}>
SCF_<{ON|OFF}>
SIGCFAIL?
Query:
SCF?
TOS9213S/ TOS9213AS