Kikusui TOS9213S Operation Manual page 214

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Indicator
DCW
IR
Testing
function
AUTO
DCW
Memory
IR
function
AUTO
Backup battery life
MEASURE MODE
NORM
MAX/MIN
PASS HOLD TIME
MOMENARY
FAIL MODE
TEST TIME
DOUBLE ACTION
KEYLOCK
10-10
240x64-dot LCD. Displays settings, measurements and judgement results.
Executes DC withstanding voltage test
Executes insulation resistance test
Automatically executes DC withstanding voltage test (DCW), insulation
resistance test (IR), and earth continuity test (EC). The TOS6200 (Earth
Continuity Tester) is necessary for earth continuity testing.
Maximum of 100
Maximum of 100
Maximum of 100 test patterns, with each having up to 100 steps (total of 500
steps)
At least 3 years (at 25 °C)
Displaying the measured current in withstanding voltage testing and the
measured insulation resistance in insulation resistance testing can be
selected as shown below
Displays, during the test, the measured current in withstanding voltage test-
ing and the measured insulation resistance in insulation resistance testing
Displays, during the test, the maximum measured current in withstanding
voltage testing and the minimum measured insulation resistance in insula-
tion resistance testing during one test
The time during which a PASS judgement is held can be set from 0.2 s to
10.0 s at a resolution of 0.1 s, or to HOLD.
Executes a test only while the START switch is pressed
Cancels to reset the FAIL or PROTECTION status produced by the STOP
signal with remote control
Starts a test only when the START switch is pressed within approximately
0.5 s after the STOP switch is pressed
Shifts to the status in which no keys other than the START and STOP
switches are accepted
TOS9213S/ TOS9213AS