Kikusui TOS9213S Operation Manual page 176

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

NOTE
PRGRETURN (PRET)
7-52
• All parameters must be entered, even if the value isn't changed.
• Do not skip a step number. Overwrite the existing step, or add a new step after the
last step.
• Up to 500 steps can be used in a set of programs.
For example, if 100 steps are set for each of programs 0 to 4, no step is available
for program 5.
(Example) At step 15 of program 10, to program memory 3 for DC withstanding
voltage testing at an interval time of 2 s,
PED 10,15,1,3,2.0
Response
To PED?_<program number, step number>, the contents of the specified
step of a specified program are returned.
(Example) If, at step 15 of program 10, memory 3 for DC withstanding voltage test-
ing is programmed at an interval time of 2 s,
"1,3,2" is returned.
Makes the RETURN/END settings for a specified program.
Also inquires about the RETURN/END setting for a specified program.
Message
Setting: PRGRETURN_<program number,{ON|OFF}>
PRET_<program number,{ON|OFF}>
PRGRETURN?_<program number>
Query:
PRET?_<program number>
Program data
Data format: Integer
Set value:
0 to 99
Resolution:
1
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set program 10 to RETURN,
PRET 10,ON
Response
To PRET?_<program number>, the RETURN setting for the specified program
is returned.
(Example) If program 10 is set to the END mode,
0 is returned.
<program number>
<program name>
TOS9213S/ TOS9213AS