(Example) To set the GND to GUARD,
Response
To I:GND?, the current GND setting is returned.
(Example) If the GND is currently set to LOW,
IR:SCAN (I:SCAN)
Sets HIGH/LOW/OPEN for the scanner channel in an insulation resistance test.
Also inquires about the current setting of the high-voltage scanner channel.
Message
Program data
Program data
(Example) To set scanner channel 1 to HIGH,
Response
To I:SCAN? 2, the scanner channel 2 setting is returned.
(Example) If the scanner channel is currently set to OPEN,
TOS9213S/ TOS9213AS
Setting: IR:GND_<{GUARD|LOW}>
I:GND_<{GUARD|LOW}>
IR:GND?
Query:
I:GND?
Data format: Character
Set value:
LOW (0), GUARD (1)
I:GND 1
0 is returned.
Setting: IR:SCAN_<channel, {H|L|O}>
I:SCAN_<channel, {H|L|O}>
IR:SCAN?_<channel>
Query:
I:SCAN?_<channel>
<channel>
Data format: Integer
Set value:
1 to 16
Resolution:
1
<{H|L|O}>
Data format: Character
Set value:
O (0): OPEN; L (1): LOW; H (2): HIGH
I:SCAN 1,H
0 is returned.