Turning the timer ON/OFF
1.00
Make ON/OFF settings for the timer function.
When the timer function is turned on, the test time can be controlled, as shown in
"Setting the test time" in the next section. When the preset test time has elapsed
with a leakage current recorded between the lower and upper references, the test
ends with a PASS judgement.
You can set the timer function to ON/OFF using the SHIFT + F5 keys, regardless of
the location of the cursor. Each time the SHIFT + F5 keys are pressed, ON and OFF
alternate.
The
function. If the cursor is at ON/OFF of the timer function, the rotary knob can be
used to make settings.
Setting the test time (TEST TIME)
1.00
The time during which a preset voltage is applied to the DUT can be set to 0.3 s
through 999 s (at a resolution of 0.1 s for the 0.3 s to 99.9s range and 1 s for the 100
s to 999s range).
When the cursor is at the timer, the rotary knob can be used to make settings.
1.
2.
3.
TOS9213S/ TOS9213AS
DCW1
DCW1
LOWER
OFF
UPPER
5.00
kV
mA
LOWER
ON/OFF
F1
F2
F3
keys can also be used to move the cursor to ON/OFF of the timer
Turn the rotary knob clockwise
Turn the rotary knob counterclockwise : OFF
DCW1
DCW1
LOWER
OFF
UPPER
5.00
kV
mA
VOLT
LOWER
UPPER
F1
F2
F3
To move the cursor to the timer, press the F5 key (TIMER). (The
keys can also be used.)
Using the
keys, move the cursor to the target digit.
Using the rotary knob, set the test time.
READY
TIMER
OFF
3.0
s
TIMER
ON/OFF
F4
F5
SHIFT
: ON
READY
TIMER
OFF
3.0
s
TIMER
F4
F5
SHIFT
5-9