Kikusui TOS9213S Operation Manual page 175

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

PRGTEST (PTES)
Sets the specified program.
Message
Program data
(Example) To set program 10,
PRGEDIT (PED)
Creates a program using stored memory. Also inquires about the contents of the pro-
gram corresponding to a program number and step number.
Following execution of this message, the program edit screen automatically appears.
Message
Program data
Program data
Program data
Program data
Program data
TOS9213S/ TOS9213AS
Setting: PRGTEST_<program number>
PTES_<program number>
Data format: Integer
Set value:
0 to 99
Resolution:
1
PTES 10
Setting: PRGEDIT_<program number, step number, mode
selection {1|2|3}>, memory number, interval
time>
PED_<program number, step number, mode selec-
tion {1|2|3}, memory number, interval time>
PRGEDIT? (PED?)_<program number, step number>
Query:
PED?_<program number, step number>
<program number>
Data format: Integer
Set value:
0 to 99
Resolution:
1
<step number>
Data format: Integer
Set value:
0 to 99
Resolution:
1
<mode selection {1|2|3}>
Data format: Integer
Set value:
1: DCW; 2: IR; 3: EC
<memory number>
Data format: Integer
Set value:
0 to 99
Resolution:
1
<interval time>
Data format: Real number, character
Set value:
0.2 to 9.9, HOLD
Resolution:
0.1
Unit:
s
7-51