Kikusui TOS9213S Operation Manual page 173

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Program data
Program data
Program data
Program data
Program data
Program data
Program data
Program data
Program data
Program data
TOS9213S/ TOS9213AS
Resolution:
0.1 E-6 (0.1 E-6 to 99.9 E-6), 1 E-6 (100 E-6 to 1000 E-6)
Unit:
A
<test time>
Data format: Real number
Set value:
0.5 to 999
Resolution:
0.1 for 0.5 to 99.9, 1 for 100 to 999
Unit:
s
<lower judgment{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<upper judgment{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<timer{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<voltage rise time>
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
<WAIT TIME>
Data format: Real number
Set value:
0.3 to 10.0
Resolution:
0.1
Unit:
s
<GND{GUARD|LOW}>
Data format: Character
Set value:
LOW (0), GUARD (1)
<contact check{ON|OFF}>
Data format: Character
Set value:
OFF, ON (0, 1)
<scanner Hi/Low>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
LOW is represented by 0 and HIGH by 1.
<scanner Open>
Data format: Integer
Set value:
0 to 65535 (#HFFFF)
Resolution:
1
OPEN is represented by 1 and NOT OPEN by 0.
OPEN has priority over HIGH/LOW.
7-49