Kikusui TOS9213S Operation Manual page 208

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Time
Setting range for the voltage
rise time (RISE TIME)
Setting range for the test time
(TEST TIME)
Accuracy
Setting range for the judge-
ment wait time (WAIT TIME)
*3 When the GND LOW/GUARD setting is set to LOW, the humidity must not exceed 70 % rh.
*4 With the GND set to GUARD, current measurement is disabled if the part of the DUT connected to the LOW terminal is
grounded, which poses extreme danger. Never ground the DUT. In ordinary operation, set the GND to LOW.
*5 In the MID and SLOW modes, depending on the discharge method, the voltage monitoring function may operate and the
TOS9213S/ TOS9213AS may enter the PROTECTION status before UPPER FAIL detection takes place.
10-4
0.1 s to 200 s
0.3 s to 999 s (With the TIMER OFF function)
± (100 ppm + 20 ms)
0.3 s to 30 s [(RISE TIME + TEST TIME) > WAIT TIME]
TOS9213S/ TOS9213AS