Kikusui TOS9213S Operation Manual page 210

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Judgement function
Judgement method/action
Resistance
Setting range
value-based
for the upper
judgment
reference
Current value-
(UPPER)
based judgment
Resistance
Setting range
value-based
for the lower
judgment
reference
Current value-
(LOWER)
based judgment
10-6
• In UPPER and LOWER judgement, you can switch between resistance value-based
judgement and current value-based judgement.
Judgement
Judgement method
When the tester detects a resistance
or a current exceeding the upper ref-
erence, it cuts off the output and
makes an UPPER FAIL judgement.
However, in resistance value-based
UPPER
judgement, judgement does not take
FAIL
place while the voltage is rising
(RISE TIME), and in current value-
based judgement, judgement does
not take place during the judgement
wait time (WAIT TIME).
When the tester detects a resistance
or a current below the lower refer-
ence, it cuts off the output and
makes a LOWER FAIL judgement.
However, in resistance value-based
LOWER
judgement, judgement does not take
FAIL
place during the judgement wait time
(WAIT TIME), and in current value-
based judgement, judgement does
not take place while the voltage is
rising (RISE TIME).
When the preset time has elapsed
without any abnormalities, the tester
PASS
cuts off the output and makes a
PASS judgement.
• The PASS signal is output at the timing preset on PASS HOLD. If HOLD is set, the
PASS signal is output continuously until the STOP signal is input.
• The UPPER FAIL signal and the LOWER FAIL signal are output continuously until the
STOP signal is input.
• The FAIL and PASS buzzer volumes are adjustable. However, they cannot be adjusted
individually, as they are set in common.
0.01 MΩ to 9.99 GΩ [Below the maximum rated current]
0.1 µA to 1.00 mA
0.01 MΩ to 9.99 GΩ [Below the maximum rated current]
0.1 µA to 1.00 mA
Display
Buzzer SIGNAL I/O
The FAIL
LED lights
Outputs
up.
ON
the U FAIL
Displayed on
signal
the LCD
The FAIL
LED lights
Outputs
up.
ON
the L FAIL
Displayed on
signal
the LCD
The PASS
LED lights
Outputs
up.
ON
the PASS
Displayed on
signal
the LCD
TOS9213S/ TOS9213AS