Response
To SIL?, the set value for an acknowledge message is returned.
(Example) If the current settings are made to "Do not return an acknowledge mes-
START
Starts testing.
While a test program is running, it shifts the program from the step suspended due
to a HOLD setting in the interval to the next step.
This command message is valid only in the test conditions set up (ACW, DCW, IR),
offset measurement (OFFSET), or program (AUTO) screens. Switch screens using
the FUNCTION message.
Message
STOP
Performs the following functions:
Message
TRM
Used to set the response-message terminator, then inquire about the set value it.
Note that the GPIB uni-line message "EOI" is effective only in GPIB communica-
tion.
Message
Program data
(Example) To set the response-message terminator to LF+EOI,
Response
To TRM?, the preset response-message terminator is returned.
(Example) If the response-message terminator is set to LF+EOI, 1 is returned.
TOS9213S/ TOS9213AS
sage," 1 is returned.
Setting: START
• Stops testing.
• Clears PASS and FAIL judgements and the PROTECTION status.
• Stops forced discharge (unless the discharge time is set to 0.0 s)
Setting: STOP
Setting: TRM_<{0|1|2|3}>
TRM?
Query:
Data format: Integer
Set value:
0: CR/LF+EOI; 1: LF+EOI; 2: EOI; 3: CR+EOI
TRM 1
7-15