Kikusui TOS9213S Operation Manual page 158

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

RDATA (RDAT)
NOTE
TIME
MON
7-34
Inquires about the monitor resistance.
During the test, the measured normal value or the minimum value is returned,
depending on the MEAS MODE setting in the system settings. After the test, the
previous monitor resistance is returned.
Message
RDATA?
Query:
RDAT?
Response
To RDAT?, the monitor resistance is returned for IR in an insulation resistance test.
(Example) If the current monitor resistance is 10.0 MΩ,
10.0E6 is returned.
• In insulation resistance testing, when the resistance value to exceed 9.99 GΩ is detected.
9.99 GΩ < 99.9 GΩ
99.9 GΩ
Inquires about the test time elapsed (the remaining time when the timer function is
on) and the test status.
Message
TIME?
Query:
Response
To TIME?, the time elapsed (the remaining time when the timer function is on) and
the test status are returned as "time, status."
Rising (Rise): 0; Testing (Test): 1; Ending (End): 3
(Example) If the time elapsed (or the remaining time) is 7.0 s and the test is under
way,
7.0,1 is returned.
Inquires about each monitor value
Message
MON?
Query:
Response
To MON?, the current monitor value is returned.
When no test is under way, the previous test results are returned.
Returning in DC withstanding voltage testing
Test type (1: DCW), monitor voltage, monitor current NORM, monitor current
MAX, monitor resistance MIN, elapsed (remaining) time, test status (0: Rise/1:
Test/3: End)
:The returned resistance value is not warranted.
: "99.9" flashes on the LCD and -1 is returned.
TOS9213S/ TOS9213AS