DCW:RISETIME (D:RTIM)
Sets the voltage rise time (RISE TIME) for a DC withstanding voltage test. Also
inquires about the set or setting voltage rise time.
Message
Program data
(Example) To set the voltage rise time to 1 s,
Response
To D:RTIM?, the current voltage rise time is returned.
(Example) If the voltage rise time is currently set to 2.0 s,
DCW:DISCH (D:DIS)
Sets the discharge time (DISCH TIME), which is the duration of the forcible dis-
charge that occurs after DC withstanding voltage testing. Also inquires about the
current discharge time.
Message
Program data
(Example) To set the discharge time to 1 s,
Response
To D:DIS?, the current discharge time is returned.
(Example) If the discharge time is currently set to 2.0 s,
TOS9213S/ TOS9213AS
Setting: DCW:RISETIME_<voltage rise time>
D:RTIM_<voltage rise time>
DCW:RISETIME?
Query:
D:RTIM?
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
D:RTIM 1
2.0 is returned.
Setting: DCW:DISCH_<DISCH TIME>
D:DIS_<DISCH TIME>
DCW:DISCH?
Query:
D:DIS?
Data format: Real number
Set value:
0.0, 0.5 to 300
Resolution:
0.1 for 0.5 to 99.9, 1 for 100 to 300
Unit:
s
D:DIS 1
2.0 is returned.