Kikusui TOS9213S Operation Manual page 56

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Setting the upper reference (UPPER)
CAUTION
NOTE
5-8
• If the upper reference is set to above 2.5 mA, the protective circuit may be
activated. To avoid this, set the output time to 1 minute or less and provide
a pause longer than the test time.
DCW1
DCW1
LOWER
OFF
UPPER
1.00
5.00
kV
VOLT
LOWER
UPPER
F1
F2
F3
The upper reference can be set to 1 µA through 5.50 mA (at a resolution of 0.01 mA
for the 1.00 mA to 5.50 mA range, and 1 µA for the 1 µA to 999 µA range).
Once a current exceeding the upper reference is detected, the test ends with a FAIL
judgement.
If the cursor is at the upper reference, the rotary knob can be used to make settings.
1.
If the upper reference is not displayed, press the F3 key (UPPER) to
display it (while the upper reference is displayed, the cursor can be
moved using the
2.
Using the
keys, move the cursor to the target digit.
3.
Using the rotary knob, set the upper reference.
• With the lower judgement function on, if the upper reference is set at or below the
lower reference, "READY" disappears and "UP<=LOW" flashes at the top right
of the LCD to indicate that testing cannot be performed (the lower reference is
factory-set to 0.20 mA). Lower the lower reference, raise the upper reference, or
turn off the lower judgement function.
• When settings are being made, if the test voltage multiplied by the upper reference
exceeds 55 W, "READY" disappears and "OVER 55 W" flashes at the top right of
the LCD to indicate that testing cannot be performed. In such a case, lower the test
voltage or the upper reference.
READY
TIMER
OFF
3.0
s
mA
TIMER
F4
F5
SHIFT
keys).
TOS9213S/ TOS9213AS