Kikusui TOS9213S Operation Manual page 141

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Program data
(Example) To set the lower reference to 1 mA,
Response
To D:LOW?, the currently set lower reference and the ON/OFF status of the lower
judgment function are returned as "lower reference, ON/OFF."
(Example) If the lower reference is currently set to 2.0 mA and the lower judgment
DCW:UPPER (D:UPP)
Sets the upper reference (UPPER) for a DC withstanding voltage test. Also inquires
about the set or setting upper reference.
CAUTION
• If the upper reference is set above 5 mA, the protective circuit may be acti-
vated. To prevent this, set the output time below 1 minute and the pause to
longer than the output time.
Message
Program data
(Example) To set the upper reference to 5 mA,
Response
To D:UPP?, the preset upper reference is returned.
(Example) If the upper reference is currently set to 2.00 mA,
TOS9213S/ TOS9213AS
Resolution:
0.01 E-3 for 1.00 E-3 to 5.50 E-3, 0.001 E-3 for 0.001 E-3 to
0.999 E-3
Unit:
A
<{ON|OFF}>
Data format: Character
Set value:
OFF (0), ON (1)
D:LOW 1E-3,ON
function to OFF,
"2.00E-3,0" is returned.
Setting: DCW:UPPER_<upper reference>
D:UPP_<upper reference>
DCW:UPPER?
Query:
D:UPP?
Data format: Real number
Set value:
0.01 E-3 to 5.50 E-3
Resolution:
0.01 E-3 for 1.00 E-3 to 5.50 E-3, 0.001 E-3 for 0.001 E-3 to
0.999 E-3
Unit:
A
A:UPP 5E-3
2.00E-3 is returned.