■ TOS6200 protection
When the TOS6200, while controlled by the tester, enters the PROTECTION status,
the tester itself also enters the PROTECTION status. "TOS6200" flashes on the
LCD.
To cancel the status, eliminate the cause of the problem with the TOS6200 and then
press the STOP switch.
01 DCW 00-1.0s
02 IR
03 EC
■ Communication error during TOS6200 control
When an error occurs in interface settings or communication cables while the tester
is controlling the TOS6200, "COM ERR" flashes on the LCD.
To cancel the protection function, eliminate the cause of the communication error
and then press the STOP switch.
01 DCW 00-1.0s
02 IR
03 EC
TOS9213S/ TOS9213AS
00 : --UNTITLED--
AUTO
AUTO
PRG
02-1.0s
03-1.0s
INS
DEL
AUTO
AUTO
00 : --UNTITLED--
PRG
02-1.0s
03-1.0s
INS
DEL
TOS6200
COM ERR
5-61