Kikusui TOS9213S Operation Manual page 119

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Control procedure
1.
2.
3.
4.
NOTE
• When the ENABLE level of the START signal is changed, "REMOTE" flashes on
the LCD. The tester enters the PROTECTION status ("PROTECTION" lights up).
To cancel the PROTECTION status, use the STOP switch on the panel or the
STOP signal.
• If FAIL MODE is set to ON in the system settings, neither the FAIL status nor the
PROTECTION status can be cancelled using the STOP signal from the remote
control. To cancel these statuses, use the STOP switch on the panel.
For the system settings, see "5.8 System Settings".
• The input terminal is pulled up to +12 V by resistance. Opening the input terminal
is equivalent to inputting a high level.
TOS9213S/ TOS9213AS
Short-circuit ENABLE of the START signal on pin 23 using pin 13 or pin
25 (COM) to shift to the low level. The START switch on the panel and
the START input of REMOTE will then become invalid. A stop operation
is workable with the STOP switch on the panel, or by inputting STOP
from the REMOTE terminal or the STOP signal from SIGNAL I/O.
While the READY signal on pin 19 is ON, short-circuit the START signal
on pin 21 using pin 13 or pin 25 (COM) to shift to the low level. The test
will then start.
Short-circuit the STOP signal on pin 22 using pin 13 or pin 25 (COM) to
shift to the low level. The test will then stop.
To cancel the remote-control function, set ENABLE of the START signal
to the high level.
The START switch on the panel will then become valid, while the START sig-
nal for SIGNAL I/O will become invalid.
6-9