Kikusui TOS9213S Operation Manual page 113

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Control procedure
1.
2.
3.
4.
5.
6.
NOTE
• When the START signal of the SIGNAL I/O connector is valid, the signal has pri-
ority over the START signal of the REMOTE terminal. The STOP signals of the
SIGNAL I/O connector and the REMOTE terminal are accepted equally.
• If the REMOTE connector is inserted/extracted with the POWER switch on,
"REMOTE" flashes on the LCD. The tester enters the PROTECTION status
("PROTECTION" lights up) and cuts off the high-voltage output.
• In making system settings, when FAIL MODE is set to ON, the FAIL or PRO-
TECTION status cannot be cancelled even if the STOP signal is input from the
REMOTE terminal. To cancel the statuses, press the STOP switch on the panel.
For the system settings, see "5.8 System Settings".
TOS9213S/ TOS9213AS
Turn off the POWER switch.
Using the exclusive connection cable (5-pin DIN cable), connect the
optional device to the REMOTE connector on the front panel.
Turn on the POWER switch.
A test can be started on the optional device. In such a case, the START switch
on the panel becomes invalid. To stop the test, either press the STOP switch on
the panel or input the STOP signal on the optional device. For details, see the
operation manual for the optional device.
To return to the panel for control, first turn off the POWER switch on the
tester.
On the front panel, remove the exclusive connection cable (5-pin DIN
cable) from the REMOTE connector.
Turn on the POWER switch.
The START switch on the panel becomes valid.
6-3