Kikusui TOS9213S Operation Manual page 44

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

[7] MEMORY
[8] LCD
[9] Analog voltmeter
WARNING
CAUTION
[10] Analog-voltmeter zero adjuster
CAUTION
[11] DANGER lamp
WARNING
4-4
AUTO/EDIT
When this key is pressed, the LED lights up. The LCD displays the program
execution screen (AUTO READY).
To display the program edit screen (AUTO EDIT), press this key together with
the SHIFT key.
RECALL/STORE
Used to recall the panel memory.
To change the memory number, first press the rotary knob, then press the
ENTER key next to the rotary knob to recall the contents.
To store data in memory, press this key together with the SHIFT key.
Displays settings and measurements.
Voltmeter used to display the output voltage. Directly reads the voltage between the
HIGH VOLTAGE terminal and the LOW terminal.
• While the pointer of the analog voltmeter is moving, never touch the HIGH
VOLTAGE terminal, test leadwire, or DUT.
• This voltmeter cannot be used as an independent voltmeter. It may mal-
function if a voltage is applied to the output terminal from outside.
Adjuster used to adjust the analog voltmeter to the zero point.
• Prior to zero adjustment, be sure to turn off the POWER switch.
Red lamp indicating that a high voltage is being output.
This lamp lights up during testing and automatic testing, or while an output voltage
remains in the output terminal.
• While this lamp is lit, never touch the HIGH VOLTAGE terminal, test lead-
wire, or DUT.
TOS9213S/ TOS9213AS