Kikusui TOS9213S Operation Manual page 183

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Fail register
Bit
7
6
5
4
3
2
UPPER FAIL
1
LOWER FAIL
0
CONTACT FAIL
Invalid-Setting register
Bit
7
6
5
4
SCANNER CHANNEL
3
WAIT TIME
2
UP
LOW
OVER 1.1mA,
1
OVER 0.11mA
(for TOS9213AS only)
0
OVER 55W
Error register
Bit
7
6
5
4
3
Invalid message
2
Out-of-range error
1
Data Error
0
Syntax Error
TOS9213S/ TOS9213AS
Not used with this tester
Not used with this tester
Not used with this tester
Not used with this tester
Not used with this tester
Indicates FAIL for the upper cutoff value
Indicates FAIL for the lower cutoff value
Indicates FAIL in a contact check
Not used with this tester
Not used with this tester
Not used with this tester
When an unconnected channel has been specified
At rise time + test time
wait time
When the lower cutoff value exceeds the upper cutoff value
OVER1.1 mA indicates that the test voltage divided by the lower refer-
ence has exceeded 1.1 mA in IR.
OVER0.11mA indicates that the test voltage divided by the lower/upper
reference has exceeded 0.11 mA in IR when the test voltage setting
exceeds 1020 V on TOS9213AS.
When the test voltage multiplied by the upper reference has exceeded 55
W in DCW
Not used with this tester
Not used with this tester
Not used with this tester
Not used with this tester
Indicates an invalid message
Indicates an out-of-range error
Indicates a data error
Indicates a header error
Description
Description
Description
7-59