Returning in insulation resistance testing
Test type (2: IR), monitor voltage, monitor resistance NORM, monitor resistance
MIN, monitor current, 0, elapsed (remaining) time, test status (0: Rise/1: Test/3: End)
(Example) If an DC withstanding voltage test is under way at a monitor voltage of
NOTE
• In insulation resistance testing with the lower judgment function off, even if set-
tings are made so that the test voltage divided by the lower reference exceeds 1.1
mA, testing can be performed. However, when the voltage value which is out of
the range of ±(10 % of setting +50 V) is detected during the test, the measurement
voltage value flashes on the LCD. In this case, -1 is returned. In this case, -1 is
returned.
• In insulation resistance testing, when the resistance value to exceed 9.99 GΩ is detected.
9.99 GΩ < 99.9 GΩ
99.9 GΩ
SCANTYPE (STYP)
Inquires about the number of valid connected channels and the model name of the
scanner.
Message
Response
To STYP?, the number of valid channels and the model name of the scanner are
returned as follows:
Number of valid channels, scanner 1, scanner 2, scanner 3, scanner 4
0: None; 1: TOS9220; 2: TOS9221
(Example) If there are 12 valid channels and scanners TOS9220, TOS9220, and
CFAILCH (CFCH)
Inquires about the channel in the C FAIL status.
Message
Response
To CFCH?, the channel in the CONTACT FAIL status is returned.
(Example) If channel 12 is currently in the CONTACT FAIL status,
TOS9213S/ TOS9213AS
2.50 kV, monitor current NORM of 5.2 mA, monitor current MAX of 12
mA, monitor resistance MIN of 10 MΩ, and remaining time of 2.6 s,
"1,2.5E3,5.2E-3,12E-3,10E6,0,2.6,1" is returned.
:The returned resistance value is not warranted.
: "99.9" flashes on the LCD and -1 is returned.
SCANTYPE?
Query:
STYP?
TOS9221 (in that order, from the nearest to the tester) are connected
without a fourth scanner connected,
"12,1,1,2,0" is returned.
CFAILCH?
Query:
CFCH?
12 is returned.