Kikusui TOS9213S Operation Manual page 152

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

IR:TIMER (I:TIM)
IR:RISETIME (I:RTIM)
7-28
Sets the test time (TEST TIME) and ON/OFF of the timer function for an insulation
resistance test. Also inquires about the current test time and the ON/OFF status of
the timer function.
Message
Setting: IR:TIMER_<test time, {ON|OFF}>
I:TIM_<test time, {ON|OFF}>
IR:TIMER?
Query:
I:TIM?
Program data
Data format: Real number
Set value:
0.5 to 999
Resolution:
0.1 for 0.5 to 99.9, 1 for 100 to 999
Unit:
s
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the test time to 5 s,
I:TIM 5,ON
Response
To I:TIM?, the current test time and the ON/OFF status of the timer function are
returned.
(Example) If the test time is currently set to 2 s and the timer function to OFF,
"2.0,0" is returned.
Sets the voltage rise time (RISE TIME) for an insulation resistance test. Also
inquires about the set or setting voltage rise time.
Message
Setting: IR:RISETIME_<voltage rise time>
I:RTIM_<voltage rise time>
IR:RISETIME?
Query:
I:RTIM?
Program data
Data format: Real number
Set value:
0.1 to 200
Resolution:
0.1 for 0.1 to 99.9, 1 for 100 to 200
Unit:
s
(Example) To set the voltage rise time to 1 s,
I:RTIM 1
Response
To I:RTIM?, the current voltage rise time is returned.
(Example) If the voltage rise time is currently set to 2.0 s,
2.0 is returned.
<test time>
<{ON|OFF}>
TOS9213S/ TOS9213AS

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tos9213as

Table of Contents