Kikusui TOS9213S Operation Manual page 182

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Device Status register, Device-Status Enable register
Bit
7
PROT (Protection)
6
STOP
5
FAIL
4
PASS
3
HV ON
2
TEST
1
INV SET (Invalid setting)
0
READY
Protection register 1
Bit
7
6
5
4
OVER LOAD
3
LVP (Low Voltage Protection)
2
OCP (24 V)
1
0
OHP (Over Heat Protection)
Protection register 2
Bit
7
TOS6200 PROT (Protection)
TOS6200 COMM
6
(Communication Error)
5
REN (Remote Enable)
4
REM IO (Remote In/Out)
3
SCAN (Scanner In/Out)
2
INTERLOCK
1
10 %±50 V (Setting 10 %±50 V)
0
100/200 (Select 100/200)
7-58
Protection status
Suspension of the test
FAIL judgment
PASS judgment
Waiting for a contact check and outputting a voltage
Outputting a test voltage
LCD displaying an invalid message such as "UP
READY status
Contact Check
TEST
HV ON
Fig.7-7
Relation of TEST (Bit 2) to HV ON (Bit 3).
Not used with this tester
Not used with this tester
Not used with this tester
When 55 W is exceeded in DCW
When 1.1 mA is exceeded in IR
When a 24 V overcurrent has been detected in the SIGNAL I/O connector
or INTERLOCK connector
Not used with this tester
When the internal temperature has increased abnormally
When the TOS6200 is in the protection status
When an error has occurred in communication with the TOS6200
Change in the Enable signal for SIGNAL I/O
When the remote-control connector has been connected/disconnected
When the scanner connector has been connected/disconnected, or when
a channel specified at a step during the AUTO test is disconnected.
In the INTERLOCK status
When the test voltage has deviated from 10% ±50 V
When a source voltage outside the preset range has been applied
Description
LOW" or "OVER 55 W"
Description
Description
TOS9213S/ TOS9213AS