Setting the WAIT TIME
NOTE
Setting the current detection response (RESPONSE)
5-12
DCW2
DCW2
START : 10%
RISE
: 0.1s(−−−V/s)
DISCH : 0.0s
WAIT
: 0.3s
START V
RISE TM
DISCH
F1
F2
F3
In DC withstanding voltage testing, if a test voltage is applied to a DUT that con-
tains capacitive elements, a large charge current may flow until charging is com-
pleted. To avoid upper fail judgement by the charge current, a wait time must be
provided from the starts of START VOLTAGE, and upper fail judgement will be
ignored during wait time.
Set the wait time to 0.3 s through 30.0 s (at a resolution of 0.1 s).
When the cursor is located to the right of WAIT, the rotary knob can be used to
make settings.
1.
To move the cursor to the right of WAIT, press the F4 key (WAIT TM).
(The
2.
Using the rotary knob, set the wait time.
• With the timer ON, if the wait time thus set exceeds the sum of the voltage rise
time and the test time, "OVER WAIT" flashes on the LCD to indicate that testing
cannot be performed.
DCW2
DCW2
START : 10%
RISE
: 0.1s(−−−V/s)
DISCH : 0.0s
WAIT
: 0.3s
RES-
GND
PONS
F1
F2
F3
By changing the integrated time constant of the current detection circuit, switch the
current detection response for UPPER FAIL judgement.
Integrated time constant
• SLOW : Approximately 40 ms
• MID
: Approximately 4 ms
• FAST
: Approximately 0.4 ms
The current detection response can be set using the SHIFT + F2 keys (RESPONSE),
regardless of the location of the cursor. Each time the SHIFT + F2 keys are pressed,
the selection is switched among SLOW, MID, and FAST, in order.
GND
: LOW
WAIT TM
F4
F5
SHIFT
keys also can be used.)
RESPONSE : SLOW
GND
: LOW
F4
F5
SHIFT
TOS9213S/ TOS9213AS