Kikusui TOS9213S Operation Manual page 72

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

Turning the upper judgement function ON/OFF
NOTE
Setting the upper reference (UPPER)
5-24
IR1
IR1
LOWER
ON
UPPER
10
UPPER
10.0
V
LOWER
UPPER
ON/OFF
ON/OFF
F1
F2
F3
The upper judgement function can be turned on/off.
When the lower judgement function is on, the test ends with a FAIL judgement if
the measured insulation resistance or the value of the leakage current that flows
through the insulation resistance drops below the lower reference.
The upper reference is explained in the following section. You can set the upper
judgement function to ON/OFF using the SHIFT + F3 keys, regardless of the loca-
tion of the cursor. Each time the SHIFT + F3 keys are pressed, ON and OFF alter-
nate.
The
keys can be used to move the cursor to ON/OFF of the upper judge-
ment function. If the cursor is at ON/OFF of the upper judgement function, the
rotary knob can be used to make settings.
Turn the rotary knob clockwise
Turn the rotary knob counterclockwise : OFF
• If the lower and upper judgement functions are off, no FAIL judgement is made.
Note that, when the timer is turned on with both functions off, a PASS judgement
is made.
You can set the lower and upper references using resistance or current values. For
details, see, "Selecting Resistance Value-based or Current Value-based Judgement"
on page 5-26.
Resistance value-based judgment
IR1
IR1
LOWER
ON
UPPER
ON
10
UPPER
10.0
V
M
VOLT
LOWER
UPPER
F1
F2
F3
F4
READY
ON
TIMER
ON
3.0
s
M
JUDGE
TIMER
I/R
ON/OFF
F4
F5
SHIFT
: ON
Current value-based judgment
READY
TIMER
ON
3.0
s
TIMER
F5
SHIFT
READY
IR1
IR1
LOWER
ON
UPPER
ON
10
UPPER
1.0
V
µA
VOLT
LOWER
UPPER
F1
F2
F3
F4
TOS9213S/ TOS9213AS
TIMER
ON
3.0
s
TIMER
F5
SHIFT

Advertisement

Table of Contents
loading

This manual is also suitable for:

Tos9213as

Table of Contents