Kikusui TOS9213S Operation Manual page 164

Tos9200 dc withstanding voltage and insulation resistance tester
Table of Contents

Advertisement

SIGTEST (STES)
SIGPASS (SPAS)
SIGUFAIL (SUF)
7-40
Sets TEST for Status Signal Output. Also inquires about the current TEST setting.
Message
Setting: SIGTEST_<{ON|OFF}>
STES_<{ON|OFF}>
SIGTEST?
Query:
STES?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set TEST for Status Signal Output to OFF,
STES OFF
Response
To STES?, the current TEST setting for Status Signal Output is returned.
(Example) If the TEST status of Status Signal Output is currently set to ON,
1 is returned.
Sets PASS for Status Signal Output. Also inquires about the current PASS setting.
Message
Setting: SIGPASS_<{ON|OFF}>
SPAS_<{ON|OFF}>
SIGPASS?
Query:
SPAS?
Program data
Data format: Character
Set value:
OFF (0), ON (1)
(Example) To set the PASS status of Status Signal Output to OFF,
SPAS OFF
Response
To SPAS?, the current PASS setting for Status Signal Output is returned.
(Example) If the PASS status of Status Signal Output is currently set to ON,
1 is returned.
Sets U FAIL for Status Signal Output. Also inquires about the current U FAIL setting.
Message
Setting: SIGUFAIL_<{ON|OFF}>
SUF_<{ON|OFF}>
SIGUFAIL?
Query:
SUF?
Program data
Data format: Character
TOS9213S/ TOS9213AS