Thermo Scientific
LogR Temperature Technology
The Versa Star Pro meter with pH/LogR measurement module allows direct temperature
measurement and temperature compensation using most standard glass-bulb pH electrodes
and patented digital LogR technology.
LogR technology is based on using the electrical resistance of the pH-sensing glass-bulb as the
temperature source. The logarithm of the resistance of the bulb varies almost linearly with the
reciprocal of the absolute temperature. Almost all standard glass-bulb pH electrodes exhibit a
similar decrease in resistance with increasing temperature. To maximize LogR performance
and accuracy, a ROSS Ultra or ROSS pH electrode is recommended.
LogR Resistance vs. Temperature Graph
LogR Temperature Accuracy
Table 1 – Two Point LogR Temperature Calibration, Average Error at pH 3 and pH 11
Average
Temp. Error
Electrode
0 to 25 °C
(°C)
8202BN
0.36
8203BN
0.13
8235BN
0.04
8256BN
0.22
8272BN
0.23
9202BN
0.21
9203BN
0.20
9206BN
0.29
9207BN
0.16
9256BN
0.27
9272BN
0.22
* For 20 °C temperature compensation spans
Appendix A
Average
Average Temperature
Temp. Error
Compensation Error
Above 20 °C *
0 to 25 ºC
(°C)
(pH Units)
0.10
0.004
0.02
0.002
0.14
0.000
0.06
0.003
0.06
0.003
0.06
0.003
0.10
0.002
0.06
0.003
0.07
0.002
0.07
0.003
0.06
0.003