Table 2-3 TTL Input Timing Diagram 1

Table 2-4 TTL Inputs Timing Diagram 2

E4419B Power Meter User's Guide
T4
Time for zero/cal operation to complete.
All timings based on 100 ms firmware polling.
Timing of zero/cal inputs for condition "00".
Inputs
Operation
Time
Description
T5
Maximum time between inputs going low.
T6
Minimum overlap of low inputs.
T5
T6
T7
Power Meter Operation
Zero Both: 10 s (each 8480
Series sensor),
12 s (each E Series sensor),
20 s (each N8480 Series sensor
exclude Option CFT),
8 s (each N8480 Series sensor
with Option CFT)
Cal: 6 s (8480 Series)
7 s (E-Series)
9 s (N8480 Series exclude
Option CFT)
6 s (N8480 Series with Option
CFT)
T8
Value
100 ms
200 ms
2
39