Making Measurements Using Frequency Dependent Offset Tables - Agilent Technologies E4419B User Manual

Power meter
Hide thumbs Also See for E4419B:
Table of Contents

Advertisement

2
Power Meter Operation
Making Measurements using Frequency Dependent Offset
Tables
This section describes how to use frequency dependent offset tables.
Frequency dependent offset tables provide you with a quick and convenient
method of compensating for your external test setup over a range of
frequencies. Note that when selected, frequency dependent offset correction is
IN ADDITION to any correction applied for sensor frequency response.
The power meter is capable of storing 10 frequency dependent offset tables of
80 frequency points each.
To use frequency dependent offset tables, you:
1 Select the table to work on a channel. Refer to
2 Zero and calibrate the power meter. The reference calibration factor used
3 Specify the frequency of the signal you want to measure. The calibration
4 Make the measurement.
56
Dependent Offset Table"
to edit the table, refer to
page 60 for further information.
during the calibration is automatically set by the power meter from the
sensor calibration table (if selected).
factor/offset is automatically set by the power meter from the sensor
calibration table (if selected) and the frequency dependent offset table.
Refer to
"Making the Measurement"
on page 57 for further information. If you require
"Editing Frequency Dependent Offset Tables"
on page 58 for further information.
E4419B Power Meter User's Guide
"Selecting a Frequency
on

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents